Winner 2005A Intelligent Laser Particle Size Analyzer
Overview: Winner2005A intelligent full automatic laser diffraction particle size analyzer with MIE scatteringtheory to precisely test the particle size distribution
from 0.01μm to 1000μm. It adopts dual laser beam and built-in sample preparation system, Intelligent operation, very simple to use for engineer, offer accurate and reliable data, built-in side wide-angle detector, make sure get 0.01 micron particles scattering angles. Main Specifications: Model Name | Winner2005A |
Standard | ISO13320-1:1999,GB/T19077.1-2008,Q/0100JWN001-2013 |
Principle | MIE scattering theory |
Size range | 0.01μm-1000μm |
Detector Channels | 90 pcs |
Accuracy error | <1% (Deviation of D50 on national standard sample) |
Repeatability error | <1% (Deviation of D50 on national standard sample) |
Laser | High performance semiconductor red laser (λ=639nm ) Auxiliary Laser: semiconductor λ= 532nm,p>2mW life time>25000 hour |
Wet dispersion | Ultrasonic | Frequency:40KHz Power:50W, Time: ≥1S |
Agitator | Revolutions Speed: 0-3000RPM (Adjustable) |
Circulation | Rated Flow:8L/min Rated Power:10W |
Sample pool | Volume: 350mL |
Anti-overflow sensor | Made in SST , UK |
Operation mode | Manual and Full automatic, 2 mode, freely choose |
Optical alignment system | Full automatic optical path alignment system |
Software function | Analysis mode | Free Distribution, R-R Distribution, Logarithm Normal Distribution, Mesh number classification etc. |
Statistic Method | Volume Distribution, Quantity Distribution |
Statistic Comparison | Several Testing Results of samples Different batches of samples testing result, Samples before and after processing, Test result of samples in different time. |
User-defined Analysis | Figure out percentage according to the particle size Figure out particle size according to the percentage Figure out percentage according to the particle size range Meet demands of representation of particle test in different industries |
Test Report | Word, Excel,Photo( Bmp), Text etc |
Multiple-language Support | Multiple language Support |
Intelligent operation | Automatically control water inflow, dispersion,test and analysis.Better Repeatability after remove human-factor |
Testing speed | <2min/time |
Outer dimension | L85cm*W39cm*H45cm |
Net weight | 40Kg |
Main Features:1)Unique Converging Light Fourier Transform Light Path, For better to test sub-micron particles, Adopt orthogonal double laser beam patented technology, measure range expanded from 0.01μm-1000μm. 2)Intelligent operation mode realize one key test, automatically supply-water,ultrasonic dispersion, circulate and drain off water, so it not only could decrease workload, but also get better repeatability after remove human-factors errors. 3) Automatic optical light alignment system is composed of precise four phase hybrid stepper motor, Its inching precision is reach to micron level,make light path be always in a good state.4) Liquid sample dispersion/preparation system: Full built-in setting mechanical stirring, ultrasonic dispersion, and circulation path in one, Big power-circulating pump and Optimization design of pipeline,effectively prevent large particle sediment.
5)
Software support: Original unconstrained free fitting patent technology make particle analysis not be restricted by any functions, truly reflect particles distribution. 6)Output parameters:(D50, D03, D06, D10, D16, D25, D75, D84, D90, D97,D98,set any value, S/V, D[3,2],D[4,3] )
Test analysis viewAfter testing, if necessary, to select records generated an average result, the system analyzes the formation of the recording. When the automatic mode test, without data processing, forming and maintaining a comprehensive system automatically analyzed test records.Adopt Patents Technology: - Optical bench design is protected by patent No.- ZL 2014 2 0378380.8,
- Three dimensional-optical bench alignment system is protected by patent No.- ZL 2013 2 0835882.4.
- MIE scattering principle application patent No.- ZL 2013 2 0812021.4.
- Dual laser beam orthogonal application is protected by patent No.-ZL 2007 2 0025702.0
- Wet circulation installation is protected by patent No.-ZL2010 2 0593526.2
Address:
Building F, North Area, University Science Industry Park, No. 750 Shunhua Road, Jinan, Shandong, China
Business Type:
Manufacturer/Factory
Business Range:
Instruments & Meters
Management System Certification:
ISO 9001
Company Introduction:
Winner Particle Instrument Stock Co., Ltd. Have been specialized in laser particle size analyzer since 1985, as the leading manufacturer in China, winner undertake the national science and technology project for many times, get more than 30 patents technology, Strongly recommended by 500 Universities in China.
Our products are including laser diffraction particle size analyzer, wet particle size analyzer, dry particle size analyzer, particle size analyzer for powders, image particle size analyzer, particle shape analyzer, spray particle size analyzer, and online particle size analyzer, apply to Pharmaceutical materials, mineral powder, metal powders, and food particle size distribution test etc. Industries.
We created more than 10 first particle size analysis technology in China. Our goal is to provide super quality and competitive price laser particle size analyzer for clients all over the world