☆ Standard Meter for AC Energy Meter Calibration Device |
Can be used as a standard meter for AC energy meter calibration devices that meet the requirements of R46. |
☆Calibrate Single /Three Phase Power Source or Meter |
|
Suitable for calibrating single/three phase standard voltage sources and standard voltmeters of class 0.05 and below (with stable voltage source). Suitable for calibrating single/three phase standard current sources and standard ammeters of class 0.05 and below (with stable current source). Suitable for calibrating single/three phase power sources, power meters, standard energy meters of class 0.05 and below (with stable power source). |
☆Comprehensive Analysis of AC power |
|
S/N | Function |
1 | The measuring channel has the function of oscilloscope, which can display the waveform of the measured power in real time. |
2 | Accurately measure the phase between the voltage and current of each phase, and visually display it through the form of phasor diagram. |
3 | Two types of harmonic distortion, THD/T (harmonic relative to full wave) and THD/F (harmonic relative to fundamental wave), are calculated. |
4 | The amplitude (RMS), content (%) and phase of the 2nd ~ 63th harmonic wave of each phase voltage or current can be analyzed in real time.; It is used to check whether the harmonic content and phase of complex waveform output of R46 device meet the requirements of the regulation. |
5 | The spectrum of each harmonic is visually displayed in the form of a bar chart (the fundamental wave is 100%). |
☆Data statistical analysis function |
|
S/N | Function |
1 | Statistical Analysis of Data: calculate the maximum value (Max), minimum value (Min), peak-peak value (P-P), average value (Avg), standard variance (S.dev), etc. |
2 | Power Stability Test: In the test cycle, real-time plot the curve of power change with time. |
3 | Normal Distribution Histogram: Displays the distribution of the collected power within a test period. |
4 | Power Stability Test: the output power stability of the inspected electric energy device can be calculated automatically. |
Power Supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz |
Maximum Power Consumption | 60 VA |
Warm-up Time | 30mins |
Temperature Performance | Working temperature: 5°C~45°C; Storage temperature: -10°C~55°C |
Humidity Performance | Operating humidity: < 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C Storage humidity: (20%~80%) R·H,no condensation. |
Altitude | < 3000 m |
Quality | Approx.10.1 kg |
Communication Interface | RS232,USB,LAN |
Size | 400 mm(W) × 345 mm(D) × 206 mm(H) (Includes handles and feet) |