Newborn Resuscitaire Neonatal Infant Radiant Warmer
Features:
Temperature Control Mode: pre-heating control, servo control & manual control;
Instant Parameter Display: set temp, skin temp, heat power rate;
Built-in APGAR Timer and >37ºC preset function
Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;
Tilting Bed and Rotary Radiant Head;
X-ray Cassette Tray under infant bed, examination light of high illuminance;
Use anti-blast quartz infrared tube as radiant heating source.
Specifications
Environment temperature |
+18ºC~+30ºC |
Environment relative humidity |
30ºC~75% |
Environment air velocity of flow |
<0.3m/s |
Power requirement |
≤1000VA |
Temperature control model |
Auto |
Temperature control range |
25ºC~37ºC |
Temperature override mode to |
N/A |
Temperature sensor accuracy |
≤0.3ºC |
Mattress temperature uniformity |
≤2.0ºC |
Temperature rise time |
≤45min |
Heat output indicate |
0~100% |
Radiant box horizon angle adjust range |
0~±90°step-less |
Mattress Tray Size |
68cm*54cm |
Mattress tilt adjust range |
0~±10°step-less |
Shelf Max.load |
2kg |
Heater infrared wavelength |
1~3μm |
Alarms |
Power failure alarm |
Sensor failure alarm |
Over temperature alarm:≤38.0ºC |