Apparatus: | Insulating Oil Breakdown Voltage Bdv Tester |
---|---|
Application: | Insulating Oil Dielectric Strength Tester |
Function: | Insulating Oil Breakdown Voltage Tester |
Device: | Oil Dielectric Breakdown Voltage Withstand Tester |
Test Standards: | IEC60156/Is6792/BS5874 Test, ASTM D1816, ASTM D877 |
Output Voltage: | AC 0V~100kv |
Customization: |
---|
Suppliers with verified business licenses
1 | Insulating oil breakdown test of user-defined |
2 | Proof test |
3 | IEC60156/IS6792/BS5874 test |
4 | ASTM D1816 standard test ASTM D877 standard test |
5 | 5 minutes standard test |
6 | Manual test |
7 | Calibration interface for customers, result print |
8 | Uploading data to PC and making word report automatically |
Technical Index:
1 | Output voltage:AC 0V~100KV |
2 | Rate of voltage rise: 0.5kv/s±5%, 2kv/s±5%, 3kv/s±5%,5kv/s±5% |
3 | Max test times setting for user-defined mode: 9 |
4 | Stir time setting for user-defined mode: 0~999s |
5 | Waiting time setting for user-defined mode: 0~999s |
6 | Test error: 3%RDG+0.3%FS |
7 | Display: 256´128 LCD |
8 | Maximum storage capacity: 100 groups of test data |
9 | Power supply : AC220V±10%, 50Hz/60Hz |
10 | Environment temperature: 0-50°C |
11 | Relative humidity ≤80%RH |
12 | Weight:41kg |
13 | Size: 620 mm´430mm´330mm |
More Insulating Oil Dielectric Strength Tester:
• GDYJ-501 Transformer Oil Breakdown Voltage Tester (80KV 100KV for choice)
• GDYJ-502A Transformer Oil Breakdown Voltage Tester (80KV 100KV for choice)
• YJJ-II Insulating Oil Dielectric Strength Tester
Suppliers with verified business licenses