Classification: | Physiological Functions of Diagnosis and Monitoring Equipment |
---|---|
Type: | Patient Monitor |
Certification: | CE, FDA, EEC, MSDS, ISO13485 |
Group: | Doctor and Patients, Adult Pediatric |
Measured Parameter: | ECG, SpO2, NIBP, Resp, Temp, Pr |
Applications: | Clinical, Amubulance, Transport, Hemodynamic |
Suppliers with verified business licenses
ECG | GRAND150P |
Input | 3/5 wire ECG cable |
Lead section | I II III aVR, aVL, aVF, V |
Sensitivity ( gain) selection | x 1/4, x 1/2, x 1, x 2, x 4 and auto |
Sweep speed | 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s |
Heart rate measuring range | 15-350bpm |
HR measuring accuracy | ±1bpm or ±1% (select the bigger data) |
Band width | 0.05-150Hz(diagnostic), 0.5-40Hz(Monitoring), 1-20Hz (Operation) |
CMRR | ≥90dB (diagnostic), ≥105dB (monitoring and operation) |
HR delay time | ≤10s |
Pacemaker pulse detection and rejection function | yes |
NIBP | |
Test method | Oscillometer |
Typical measuring time | <30 seconds ( typical adult cuff) |
Philosophy | Adult, Pediatric and Neonate |
Measurement range |
Systolic pressure, diastolic pressure, mean arterial pressure |
Systolic pressure | Adult: 40-275 mmHg Pediatric: 40-200 mmHg Neonatal: 40-135 mmHg |
Diastolic pressure | Adult: 10-210 mmHg Pediatric: 10-150 mmHg Neonatal: 10-95 mmHg |
Mean arterial pressure | Adult: 20-230 mmHg Pediatric: 20-165 mmHg Neonatal: 20-110 mmHg |
Measurement type | Systolic Diastolic Mean |
NIBP measuring accuracy | Mean difference ±5% mmHg, standard deviation :8mmHg |
Measurement parameter | Automatic, Continuous measurement |
Measurement method | Manual, Auto, STAT, multi-cycle mode |
Automatic measuring intervals | 1-480min |
SPO2 | |
Technique | Dual wavelength optical method |
Display Type | Waveform, Data |
Measurement range | 0-100% |
SPO2 measuring accuracy | Arms is not greater than 2% for range from 70% to 100% |
Low perfusion performance | As low as 0.3% |
Pulse rate range | 30-250bpm |
Accuracy | ±1bpm or ±2% (select the bigger data) |
Resolution | 1bpm |
Refresh cycle | 1s-2s |
IBP | |
Technique | Strain gauge transducer |
Input sensitivity | 5uV/V / mmHg |
Measuring range | -50-300mmHg |
Measuring accuracy | ±4mmHg or ±2%, whichever is greater |
Measuring positions | ART,PAP,PA, LAP, CVP, ICP, AUXPI, AUXP2 |
Calibration | zero |
Respiration (Impedance & Nasal Tube) | |
Measurement type | 0-120rpm |
Accuracy | ±1bm or ±5%, select the bigger data |
Resolution | 1rpm |
EtCO2 (Optional ) | |
Technique | Infrared optical method |
Sampling mode | Sidestream or mainstream |
Measurement range | 0-150 mmHg ±10%/min (sidestream) |
Flow rate | 50ml/min |
Measuring accuracy | 0-40 mmHg, ±2mmHg 41-70 mmHg, ±5% of reading 71-100 mmHg, ±8% of reading 101-150 mmHg, ±10% of reading |
Cardiac output (C.O.) | |
Blood temperature measuring range | 23-43ºC, accuracy ±0.5ºC |
Injecta temperature measuring range | 0-20ºC, accuracy ±0.5ºC |
Measuring range | 0.2-20L/min |
Measuring accuracy | ±0.2L/min or ±10%, whichever is greater |
Display | 15 inch TFT display |
Alarming mode | 3 level Audible-visual alarm |
Networking port | Ethernet port |