• 12.1" Color TFT Touch LCD Screen Multi-Parameter Portable Patient Monitor
  • 12.1" Color TFT Touch LCD Screen Multi-Parameter Portable Patient Monitor
  • 12.1" Color TFT Touch LCD Screen Multi-Parameter Portable Patient Monitor

12.1" Color TFT Touch LCD Screen Multi-Parameter Portable Patient Monitor

Sterilization: Non Sterilization
Feature: Reusable
Application: Clinical
Material: Plastic
Power Type: Electric
Group: Adult
Customization:
Diamond Member Since 2015

Suppliers with verified business licenses

Basic Info.

Model NO.
MCS-PM01
Screen
12.1 Inches TFT Screen of LCD
Transport Package
Standard for Export
Trademark
Mecan
Origin
China
HS Code
90181930
Production Capacity
in Stock

Product Description

12.1" color TFT touch LCD screen Multi-Parameter Portable Patient Monitor

 
Technical Specifications
 
Lead Mode                                                 5 Leads (R, L, F, N, C or RA, LA, LL, RL,V)
Lead selection                                            I, II, III, avR, avL, avF, V,

Wavef orm                                                  2 c h
Lead mode                                                  3 Leads ( R, L, F or RA, LA, LL)
Lead selection                                            I, II, III,

Wavef orm                                                  1 c h
Gain                                                             2.5mm/mV, 5.0mm/mV, 10mm/mV,
                                                                     20mm/mV, auto
HR and Alarm Range                               Adult    15 ~ 300 bpm
                                                                    Neo/Ped    15 ~ 350 bpm
Accuracy                                                     1% or      1bpm, which is greater
Resolution                                                  1 bpm
Sensitivity                                                  > 200 (uVp-p)
Differential Input Impedance              > 5 M

CMRR                                                        Monitor         > 105 dB
                                                                   Operation > 105 dB
                                                                   Diagnosis            > 85 dB

Electrode offset potential                    300mV
Leakage Current                                     < 10 uA

Baseline Recovery                                  < 3 S After Defi.
ECG Signal Range                                    8 m V ( Vp-p )
Bandwidth                                                Surgery     1 ~ 15 Hz

                                                                   Monitor     0.5 ~ 35 Hz
                                                                   Diagnostic     0.05 ~ 100 Hz
Calibration Signal                                   1 (mVp-p),  Accuracy     5%
ST Segment Monitoring Range     Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting                       Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,BRADY, MISSED BEATS,
                                          PNP, PNC,Alarm Available,Review Available

Respiration

Method                               Impedance between R-F(RA-LL)
Differential Input Impedance       > 2.5 MΩ
Measuring Impedance Range          0.3~3 Ω
Base line Impedance Range           200~4 KΩ
Bandwidth                          0.1~2.5Hz
Resp. Rate                         Measuring and Alarm Range
                                    Adult 0 ~ 120 Brpm 
                                    Neo/Ped 0 ~ 150 Brpm 
                                    Resolution 1 Brpm 
                                    Accuracy ±2 Brpm

Apnea Alarm                        10 ~ 40 S
 
NIBP 

Method                            Oscillometric
Mode                              Manual, Auto, STAT                        
Measuring Interval in AUTO Mode      1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min)
Measuring Period in STAT Mode     5 Min
Pulse Rate Range              40 ~ 240 bpm
Alarm Type                     SYS, DIA, MEAN

Measuring and alarm range        Adult Mode
                                    SYS 40 ~ 270 mmHg 
                                    DIA 10 ~ 215 mmHg 
                                    MEAN 20 ~ 235 mmHg
                                Pediatric Mode
                                    SYS 40 ~ 200 mmHg 
                                    DIA 10 ~ 150 mmHg 
                                    MEAN 20 ~ 165 mmHg
                                Neonatal Mode
                                    SYS 40 ~ 135 mmHg 
                                    DIA 10 ~ 100 mmHg 
                                    MEAN 20 ~ 110 mmHg
Resolution                      1mmHg

Accuracy                        Maximum Mean error ±5mmHg
                                Maximum Standard deviation ±8mmHg
Overpressure Protection          Adult Mode 297±3 mmHg 
                                 Pediatric Mode 240±3 mmHg 
                                 Neonatal Mode 147±3 mmHg

SpO2
Measuring Range                  0 ~ 100 % 
Alarm Range                      0 ~ 100 % 
Resolution                       1 %
Accuracy                         70% ~ 100%  2 %
                                 0%  ~  69% unspecified
Actualization interval about      1 Sec. 
Alarm Delay                      10 Sec.
Pulse Rate                       Measuring and Alarm Range 20~300 bpm
                                 Resolution       1bpm
                                 Accuracy ±3bpm
                                   
TEMP

Channel                          2
Measuring and Alarm Range    0 ~ 50 ºC
Resolution                      0.1 ºC
Accuracy                    ±0.2 ºC
Actualization intervalabout     1 Sec.
Average Time Constant           < 10 Sec.

IBP

Label                           ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range       ART 0 ~ 300 mmHg 
                                PA -6 ~ 120 mmHg
                                CVP/RAP/LAP/ICP-10 ~ 40 mmHg
                                P1 / P2-10 ~ 300 mmHg 


Press Sensor                    Sensitivity 5  uV/V/mmHg
                                Impedance 300-3000Ω

Resolution                      1 mmHg
Accuracy                        ±2% or 1mmHg which is greater
Actualization interval            about 1 secretary

Standard:  ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature
Optional:  Nellcor SpO2, EtCO2, IBP,TEMP-2, Thermal Recorder,Wall Mount,Trolly,Central station



12.1&quot; Color TFT Touch LCD Screen Multi-Parameter Portable Patient Monitor

 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now