SN-DR3 Analog Mammography
Usage:
Detect galactocele accurately. Detect intraductal papilloma. Detect mammary duct ectasia and acute mastitis. Detect proliferation of mammary gland and breast cancer.
Features:
1. Standalone System (Generator Integrated within the main machine)
2. Generator: 40 kHz, 20 kV to 40 kV in 1 kV
3. High performance X-ray tube: Molybdenum Anode,
Focal Spot Size: 0.1 mm (Small)/ 0.3 mm (Large)
4. Support ISO center rotation function
C-arm Rotation range: +180°/-135°
5. Height range of breast support surface: 750mm-1330mm (±2%)
6. C-arm: Motorized Movement, Dual side digital display
7. Compression Paddles: Dual Smart Compression Mechanism
8. Spatial Resolution: ≥15Lp/mm
9. Film size: 18*24cm or 24*30cm
10. Focal distance : 650mm
11. Additional filter: 0.03mm Molybdenum / 0.5mm Aluminum
Additional filter Switch: Automatic
12. Field adjustment: Auto/Manual
13. Remote control panel: 7inches
14. Magnification ratio: (1.8)/1
15. Exposure mode: Manual / AEC
16. Power supply: single phase, AC 220~230 V±10%
Frequency: 50/60 Hz±1 Hz
17. Max. power : 4000 VA
18. N.W.: 240 Kg
19. Packing Size: 115 X 85 X 135 CM
Specifications:
X-ray Generator
Generator Type: High Frequency Inverter 80kHz
Input Power: Single phase 220VAC, 50/60Hz
Radiographic Ratings :Large Focal Point 20-35kV/10-510mAs
Small Focal Point 20-35kV/10-100mAs
Power Rating: 6kW
X-ray Tube
Model: China Hangzhou LR01
Focal Spot Size: Dual Focus 0.2 / 0.4mm
Target Material: Molybdenum (Mo)
Port Material: Beryllium (Be)
Anode Heat Storage: 100KJ
Anode Cooling: Air cooling
Filtration: Mo(0.03mm),Al(0.5mm)
Radiographic Stand
C-ARM Vertical Movement:580mm
Rotations Degree:+175°~ -180°
SID: 650mm
Cassette Image Receptor
18×24cm
Line Voltage
220Vac@25A,Single phase
Weight
240kg